A Letter To LifeScan Customers: URGENT: MEDICAL DEVICE RECALL- OneTouch® Ultra® Meter January 20, 2006 If you use, recommend, distribute or sell OneTouch® Ultra® Blood Glucose Meters, we want to make you aware of an issue: Beginning on April 18, 2005, OneTouch Ultra Meters were redesigned to not allow users to change the unit of measure setting. In the U.S., these meters were set to the mg/dL unit of measure. However, due to a manufacturing issue, an extremely small number of these meters do allow users to change their unit of measure setting between mg/dL and mmol/L. When setting the date and time on such meters, it is possible for the user to accidentally change the unit of measure. If a meter is set to the wrong unit of measure, the user could misunderstand his or her blood glucose results. What You Should Do Until you receive your replacement, you should continue to test your blood glucose with your current OneTouch Ultra Meter. There is no issue with the accuracy of the meter. Always check the unit of measure each time you test to make sure it is set to "mg/dL." How To Contact LifeScan
|
|||||||||||||||||||||||||||||||||||||
|
|
||||||||||||||||||||||||||||||||||||||
|